Theoretical tumor edge detection technique using multiple Bragg peak decomposition in carbon ion therapy

Marta F Dias,Charles-Antoine Collins-Fekete,Guido Baroni,Marco Riboldi,Joao Seco
DOI: https://doi.org/10.1088/2057-1976/ab18e6
2019-10-18
Abstract:Purpose: The range precision in carbon ion therapy is extremely sensitive to tissue density variations. A high energy carbon beam, after crossing a high-gradient edge parallel to the beam direction, suffers from range mixing leading to the detection of multiple Bragg peaks (BPs) varying intensity and water equivalent thickness (WET). The purpose of this work was to introduce a model that determines the position of a high-gradient edge based on information acquired from carbon transmission imaging. Methods: A model was derived to determine the lateral distance between the irradiation beam propagation axis and the edge position. To validate it, carbon beams were simulated and propagated through two parametric phantoms: (1) a bone cube in a water tank and (2) a semi-cylindrical bone insert in a water tank. The method was tested in a lung tumor case where range mixing led to more than two BPs being detected, requiring an iterative BP decomposition to determine the fraction of carbon ions crossing the materials surrounding the edge of interest. Results: The theoretical model predicted the edge position relative to the beam position with an error ≤1 mm for all studied cases with a maximum dose delivered of 24 μGy. Conclusions: The method presented here is a proof of principle. It does not take into account clinical uncertainties. However, this approach provided promising results suggesting that future extension to assess the impact of clinical uncertainties should be performed.
What problem does this paper attempt to address?
The problem that this paper attempts to solve is how to accurately detect the position of the tumor edge in carbon - ion therapy. Specifically, when the carbon - ion beam passes through a high - gradient edge (such as the boundary between the tumor and normal tissue), range mixing occurs, resulting in the detection of multiple Bragg peaks (BPs). The intensities and water - equivalent thicknesses (WET) of these peaks will be different. Therefore, the paper proposes a technique based on multi - Bragg - peak decomposition for theoretical tumor - edge detection. ### Problem Background 1. **Characteristics of Carbon - Ion Therapy** - The carbon - ion beam has a limited range and forms the so - called Bragg peak (BP) at the end of its range, which is the maximum of energy deposition. - The range of the carbon - ion beam depends on the density, atomic composition, and ionization potential of the tissue. 2. **Clinical Challenges** - In the clinical environment, the carbon - ion beam is usually delivered using a pencil beam, and its lateral distribution is Gaussian. - When the carbon - ion beam passes through materials with different densities, it will result in different ranges, especially in the lung area, where there is a significant difference between the tumor density (about 0.9 g/cm³) and the lung tissue density (0.3 - 0.45 g/cm³). - These differences may lead to geographical misalignment or high - dose deposition in organs at risk (OARs). ### Purpose of the Paper The main purpose of the paper is to introduce a model that can determine the position of a high - gradient edge based on the information obtained from carbon - transmission imaging. Specifically: - **Objective**: Determine the edge position by analyzing multiple Bragg peaks generated after the carbon - ion beam passes through a high - gradient edge. - **Method**: Develop a theoretical model to calculate the lateral distance \( Y \) between the high - gradient edge and the propagation axis of the carbon - ion beam. - **Verification**: Verify the effectiveness of this model through simulation experiments, including Monte Carlo simulations using parameterized and voxelized phantoms. ### Main Contributions 1. **Theoretical Model**: Propose a method based on multi - Bragg - peak decomposition, which can accurately measure the tumor - edge position with an error of ≤1 mm. 2. **Low - Dose Advantage**: This method can obtain accurate tumor - position information with only a very small number of irradiation points, thereby minimizing the dose received by the patient (24 µGy). 3. **Clinical Application Prospects**: This method can be applied before treatment to achieve adaptive treatment - plan adjustment and reduce the dose to organs at risk. In conclusion, this paper aims to solve the problem of range uncertainty caused by tissue - density changes in carbon - ion therapy and provides a new method for accurately detecting the tumor edge.