Accelerometric patch probe for cuffless blood pressure evaluation from carotid local pulse wave velocity: design, development, and in vivo experimental study

R Arathy,P M Nabeel,Jayaraj Joseph,Mohanasankar Sivaprakasam
DOI: https://doi.org/10.1088/2057-1976/ab1a82
2019-05-28
Abstract:Objective : In this work, we demonstrate an accelerometric patch probe and associated measurementsystem for local pulse wave velocity (PWV) measurement and its application in cuffless bloodpressure (BP) measurement. Approach: The proposed system consists of dual accelerometric patch probeto capture acceleration plethysmography (APG) signals from the carotid artery. The probe wasintegrated to an application-specific analog front-end circuitry with negligible inter-channel delayand a data acquisition module. Real-time signal processing and local PWV evaluation were performedusing custom software. The functionality of the developed system and the relationship between localPWV and reference BP parameters were experimentally validated by multiple in vivo studies on acohort of 26 subjects. Inter- and intra-subject BP-local pulse transit time (PTT) models weredeveloped and used for cuffless BP measurement. Further, the reliability of the proposed methodin...
What problem does this paper attempt to address?