Development of a tyrosinase-based biosensor for bisphenol A detection using gold leaf–like microstructures

Filomeno A. D. Inroga,Manoelly O. Rocha,Vladimir Lavayen,Jacqueline Arguello
DOI: https://doi.org/10.1007/s10008-019-04252-2
IF: 2.747
2019-04-17
Journal of Solid State Electrochemistry
Abstract:A simple one-step electrodeposition without template allowed the synthesis of gold microstructures on a screen-printed carbon electrode. Chloroauric ions were reduced by applying a constant potential of − 0.6 V during 600 s. A preferential growth along the &lt;111&gt; directions produced leaf-like structures as confirmed by scanning electron microscopy and X-ray diffraction. The as-prepared Au microstructures worked as a support for tyrosinase immobilization allowing the preparation of a highly selective and sensitive biosensor for bisphenol A (BPA) detection. The cyclic voltammograms exhibit a well-defined anodic peak at 0.24 V in phosphate buffer solution (0.1 mol L<sup>−1</sup>, pH 7.0). The enzyme creates favorable conditions for the adsorption of BPA, and after 10 min of accumulation time, the calibration curve was linear in the range of 0.5–50 μmol L<sup>−1</sup> with a detection limit of 77 nmol L<sup>−1</sup> (S/<em>N</em> = 3) and a relative standard deviation (RSD%) of 0.54% (<em>n</em> = 10). Furthermore, the proposed biosensor displayed long-term stability and was successfully applied to determine BPA in spiked water samples.
electrochemistry
What problem does this paper attempt to address?