Development of an anatomically correct mouse phantom for dosimetry measurement in small animal radiotherapy research

George Soultanidis,Anna Subiel,Isaline Renard,Anna Merle Reinhart,Victoria L Green,Uwe Oelfke,Stephen J Archibald,John Greenman,Amanda Tulk,Adrian Walker,Giuseppe Schettino,Christopher J Cawthorne
DOI: https://doi.org/10.1088/1361-6560/ab215b
IF: 3.5
2019-06-21
Physics in Medicine and Biology
Abstract:Significant improvements in radiotherapy are likely to come from biological rather than technical optimization, for example increasing tumour radiosensitivity via combination with targeted therapies. Such paradigms must first be evaluated in preclinical models for efficacy, and recent advances in small animal radiotherapy research platforms allow advanced irradiation protocols, similar to those used clinically, to be carried out in orthotopic models. Dose assessment in such systems is complex however, and a lack of established tools and methodologies for traceable and accurate dosimetry is currently limiting the capabilities of such platforms and slowing the clinical uptake of new approaches . Here we report the creation of an anatomically correct phantom, fabricated from materials with tissue-equivalent electron density, into which dosimetry detectors can be incorporated for measurement as part of quality control (QC). The phantom also allows training in preclincial radiotherapy planning and cross-institution validation of dose delivery protocols for small animal radiotherapy platforms without the need to sacrifice animals, with high reproducibility.Mouse CT data was acquired and segmented into soft tissue, bone and lung. The skeleton was fabricated using 3D priniting, whilst lung was created using computer numerical control (CNC) milling. Skeleton and lung were then set into a surface-rendered mould and soft tissue material added to create a whole-body phantom. Materials for fabrication were characterized for atomic composition and attenuation for X-ray energies typically found in small animal irradiators. Finally cores were CNC milled to allow intracranial incorporation of bespoke detectors (alanine pellets) for dosimetry measurement.
engineering, biomedical,radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?
The problem that this paper attempts to solve is: in small - animal radiotherapy research, there is a lack of precise and traceable dose - measurement tools and methods, which limits the capabilities of these platforms and hinders the translation of new treatment methods into clinical applications. Specifically, the author points out that although pre - clinical radiotherapy systems (such as SARRP) can irradiate in - vivo cancer models using clinically relevant techniques, dose evaluation is very complicated in these systems. Currently, there is a lack of standardized tools and methods for traceable and accurate dose measurement, which limits the capabilities of these platforms and slows down the clinical application of new methods. To solve this problem, the author has developed an anatomically correct mouse phantom for dose measurement in small - animal radiotherapy research. This phantom is made of materials with tissue - equivalent electron density, and dosimeters can be incorporated into it for measurement as part of quality control. In addition, this phantom also allows pre - clinical radiotherapy plan training without sacrificing animals and cross - institutional verification of dose - delivery protocols for small - animal radiotherapy platforms, and is highly reproducible. In summary, this study aims to improve the dose - evaluation method in small - animal radiotherapy research by creating an anatomically correct mouse phantom that can be used for dose measurement, thereby promoting the development and clinical translation of new treatment methods.