An Approach to Rapid Determination of Tween-80 for the Quality Control of Traditional Chinese Medicine Injection by Partial Least Squares Regression in Near-Infrared Spectral Modeling

Jin-Fang Ma,Tian-Ling Chen,Xiang-Dong Zhang,Xue Xiao,Fa-Huan Ge
DOI: https://doi.org/10.1155/2019/1521035
2019-08-22
Journal of Spectroscopy
Abstract:This study established an approach to rapidly determine the Tween-80 in traditional Chinese medicine (TCM) injection by using near-infrared (NIR) spectroscopy. Totally 133 standard solutions of Tween-80 were prepared and randomly divided into calibration set and validation set, containing 109 and 24 samples, respectively. Spectral data were preprocessed and then subjected to establish a predictive model using partial least-squares (PLS). The standard error of cross validation (SECV), standard deviation of calibration (SEC), and the determination coefficient ( R ) of the established model were 0.0561, 0.0526, and 0.9986, respectively. The model was successfully applied to determine Tween-80 contents in 25 XBJ Injections and 40 FFSX Injections , and it produced satisfactory quantitative analysis results with average relative deviations 0.60% and 0.16%, respectively, for 25 XBJ Injections and 40 FFSX Injections , and the maximum relative deviations 8.57% and 7.60%, respectively. This work shows that NIR model displayed quite good predictive ability for Tween-80 quantitative analysis, which could potentially be applied to rapid determination of Tween-80 in the production process of TCM injections and other TCM products.
spectroscopy,biochemical research methods