Two-center comparison of 10 fully-automated commercial procalcitonin (PCT) immunoassays

Giuseppe Lippi,Gian Luca Salvagno,Matteo Gelati,Mairi Pucci,Claudia Lo Cascio,Davide Demonte,Diego Faggian,Mario Plebani
DOI: https://doi.org/10.1515/cclm-2019-0888
2019-09-20
Clinical Chemistry and Laboratory Medicine (CCLM)
Abstract:Abstract Background This two-center study was designed to verify comparability of procalcitonin (PCT) values among 10 different commercial immunoassays. Methods A total number of 176 routine lithium-heparin plasma samples were divided in identical aliquots and simultaneously analyzed with 10 different PCT immunoassays, including Kryptor BRAHMS PCT sensitive, Abbott Architect BRAHMS PCT, Beckman Coulter Access PCT (on Access and DXI), BioMérieux Vidas BRAHMS PCT, Diasorin Liaison BRAHMS PCT, Fujirebio Lumipulse G BRAHMS PCT, Roche BRAHMS PCT (on Cobas E801), Diazyme PCT (on Roche Cobas C702) and SNIBE Maglumi PCT. Results Highly significant correlation was always found across multiple comparisons, with correlation coefficients comprised between 0.918 and 0.997 (all p 80% (all p < 0.001), but varied largely across multiple comparisons, ranging between 90%–99% at 0.1 μg/L, 81%–99% at 0.25 μg/L, 83%–100% at 0.5 μg/L, 94%–100% at 2.0 μg/L and 90%–99% at 10 μg/L, respectively. The larger disagreement was observed comparing Diazyme PCT and Maglumi PCT with the other methods. Conclusions Although we found acceptable correlation among 10 commercial PCT immunoassays, the limited agreement at clinical decision thresholds remains a major issue, especially at lower end of PCT concentration, thus potentially contributing to jeopardize the clinical value of this biomarker.