Complementary Metal–Oxide–Semiconductor Potentiometric Field-Effect Transistor Array Platform Using Sensor Learning for Multi-ion Imaging

Nicolas Moser,Chi Leng Leong,Yuanqi Hu,Chiara Cicatiello,Sally Gowers,Martyn Boutelle,Pantelis Georgiou
DOI: https://doi.org/10.1021/acs.analchem.9b05836
IF: 7.4
2020-03-06
Analytical Chemistry
Abstract:This work describes an array of 1024 ion-sensitive field-effect transistors (ISFETs) using sensor-learning techniques to perform multi-ion imaging for concurrent detection of potassium, sodium, calcium, and hydrogen. Analyte-specific ionophore membranes are deposited on the surface of the ISFET array chip, yielding pixels with quasi-Nernstian sensitivity to K<sup>+</sup>, Na<sup>+</sup>, or Ca<sup>2+</sup>. Uncoated pixels display pH sensitivity from the standard Si<sub>3</sub>N<sub>4</sub> passivation layer. The platform is then trained by inducing a change in single-ion concentration and measuring the responses of all pixels. Sensor learning relies on offline training algorithms including <i>k</i>-means clustering and density-based spatial clustering of applications with noise to yield membrane mapping and sensitivity of each pixel to target electrolytes. We demonstrate multi-ion imaging with an average error of 3.7% (K<sup>+</sup>), 4.6% (Na<sup>+</sup>), and 1.8% (pH) for each ion, respectively, while Ca<sup>2+</sup> incurs a larger error of 24.2% and hence is included to demonstrate versatility. We validate the platform with a brain dialysate fluid sample and demonstrate reading by comparing with a gold-standard spectrometry technique.
chemistry, analytical
What problem does this paper attempt to address?