Estimation of viscoelasticity of radial artery via simultaneous measurement of changes in pressure and diameter using a single ultrasound probe

Takumi Saito,Shohei Mori,Mototaka Arakawa,Shigeo Ohba,Kazuto Kobayashi,Hiroshi Kanai
DOI: https://doi.org/10.35848/1347-4065/ab7f1c
IF: 1.5
2020-04-01
Japanese Journal of Applied Physics
Abstract:To establish an evaluation index for vascular endothelial function, we developed an ultrasonic probe that can measure changes in blood pressure and blood vessel diameter at the same position in the radial artery. Based on phantom experiments, the pressure waveforms measured using the piezoelectric effect of the ultrasonic probe element and those using a pressure sensor exhibited a high correlation (correlation coefficient: r = 0.979 – 0.999). We confirmed the continuous measurement of the relationship between changes in blood pressure and the diameter from the in vivo experiments. We could then estimate the changes in viscoelasticity by calibrating the output from the probe element to the absolute blood pressure values in advance. The average coefficients of variations were 0.02 and 0.24 for elasticity and viscosity, respectively. This study demonstrated the possibility of measuring changes in the viscoelastic moduli of the radial arterial wall due to flow-mediated dilation using the developed ultrasonic probe.
physics, applied
What problem does this paper attempt to address?