Multiple-centre Clinical Evaluation of an Ultrafast Single-Tube Assay for SARS-CoV-2 RNA
J. Wang,K. Cai,X. He,X. Shen,J. Liu,J. Xu,F. Qiu,W. Lei,L. Cui,Y. Ge,T. Wu,Y. Zhang,H. Yan,Y. Chen,J. Yu,X. Ma,H. Shi,R. Zhang,X. Li,Y. Gao,P. Niu,W. Tan,G. Wu,Y. Jiang,W. Xu
DOI: https://doi.org/10.1016/j.cmi.2020.05.007
IF: 13.31
2020-01-01
Clinical Microbiology and Infection
Abstract:Objective: To evaluate the performance of an ultrafast single-tube nucleic acid isothermal amplification detection assay for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) RNA using clinical samples from multiple centres. Methods: A reverse transcription recombinase-aided amplification (RT-RAA) assay for SARS-CoV-2 was conducted within 15 minutes at 39 degrees C with portable instruments after addition of extracted RNA. The clinical performance of RT-RAA assay was evaluated using 947 clinical samples from five institutions in four regions of China; approved commercial fluorescence quantitative real-time PCR (qRT-PCR) kits were used for parallel detection. The sensitivity and specificity of RT-RAA were compared and analysed. Results: The RT-RAA test results of 926 samples were consistent with those of qRT-PCR (330 were positive, 596 negative); 21 results were inconsistent. The sensitivity and specificity of RT-RAA was 97.63% (330/338, 95% confidence interval (CI) 95.21 to 98.90) and 97.87% (596/609, 95% CI 96.28 to 98.81) respectively. The positive and negative predictive values were 96.21% (330/343, 95% CI 93.45 to 97.88) and 98.68% (596/604, 95% CI 97.30 to 99.38) respectively. The total coincidence rate was 97.78% (926/947, 95% CI 96.80 to 98.70), and the kappa was 0.952 (p < 0.05). Conclusions: With comparable sensitivity and specificity to the commercial qRT-PCR kits, RT-RAA assay for SARS-CoV-2 exhibited the distinctive advantages of simplicity and rapidity in terms of operation and turnaround time. (C) 2020 European Society of Clinical Microbiology and Infectious Diseases. Published by Elsevier Ltd. All rights reserved.