A RAPID COMBUSTION METHOD FOR THE ACCURATE DETERMINATION OF LOW LEVELS OF TRITIUM IN BIOLOGICAL SAMPLES

Per Gerde,Tony Stephens,Alan R. Dahl
DOI: https://doi.org/10.1080/105172398243023
1998-01-01
Toxicology Methods
Abstract:The authors have developed a simple, accurate method to combust low-level, tritium-containing tissue samples by modifying a tube furnace with two combustion zones. The sample is pyrolyzed upstream under a slow stream of argon, after which oxygen is added to the gas stream,and pyrolysis gases arecombusted over a platinum catalyst. The charred sample is combusted by changing the gas from argon to pure oxygen. A small volume of steam is then passed through the tube furnace to purge adsorbed tritiated water from the walls of the apparatus. The original design was changed in two ways. (1) Gas inlets were arranged so that baffles or porous plugs are not needed to maintain stable combustion. This change greatly reduces the residual radioactivity carried over from one sample to the next (typically 0.3 0.2 %). (2) Combustion gases are completely condensed in a cold trap with liquid nitrogen so that 100% of the tritium in the sample is trapped. The colorless, trapped water is directly dissolved in scintillation fluid, giving excellent counting efficiencies. Recoveries of activity were 100% pipetting errors in blood spiked with tritiated benzo\[a]pyrene. Using this method, tissue tritium levels 50% above counting background were accurately quantitated in 0.1-2.0 g samples of blood, liver, fat, and other tissues.