Total antimony analysis by hydride generation-microwave plasma-atomic emission spectroscopy with applications

Junko Fujihara,Naoki Nishimoto
DOI: https://doi.org/10.1016/j.microc.2020.104992
IF: 5.304
2020-09-01
Microchemical Journal
Abstract:Antimony is widely used in industrial applications. In this study, we develop a method for total antimony analysis using a hydride generation-microwave plasma-atomic emission spectroscopy (HG-MP-AES) system equipped with a multimode sample introduction system (MSIS). Before analysis, antimony was reduced with potassium iodide and acidified with hydrochloric acid. The samples and a sodium tetrahydroborate/sodium hydroxide solution were infused into a spray chamber for hydride generation, MSIS. The limits of detection and quantification were 0.05 μg/L and 0.15 μg/L, respectively. Furthermore, semiconductor materials and blood samples have been analyzed in order to demonstrate the possible applications of this method. The total antimony content eluted from GaSb thin films in a pH 5 buffer agreed with the thin film structure. The spiked blood sample analysis showed that the recovery rate of antimony from the whole blood samples was 92.2 ± 1.67 %. Relative standard deviations for inter-day and intra-day assays of whole blood were 1.99 % and 5.31 %, respectively. The developed method was determined to have good accuracy and precision with the obtained values, and can be utilized for antimony analysis in the engineering and toxicological fields.
chemistry, analytical
What problem does this paper attempt to address?