Implementation of passive compliance training on a parallel ankle rehabilitation robot to enhance safety

Jianfeng Li,Wenpei Fan,Mingjie Dong,Xi Rong
DOI: https://doi.org/10.1108/ir-02-2020-0040
2020-05-11
Abstract:Purpose The purpose of this paper is to implement a passive compliance training strategy for our newly designed 2-UPS/RRR parallel ankle rehabilitation robot (PARR) to enhance its rehabilitation training safety. Design/methodology/approach First, a kinematic analysis of the PARR is introduced, and the mechanism ensures that the rotation centre of the ankle joint complex (AJC) coincides with robot’s rotation centre. Then, a passive compliance training strategy based on admittance control is described in detail and is implemented on our PARR. Findings Experiments involving healthy subjects were conducted, and the performance of trajectory tracking was quantitatively evaluated, with the results showing excellent compliance and trajectory tracking accuracy, which can ensure that a secondary injury to the AJC during passive rehabilitation training is avoided. The influence of different admittance parameters was also simulated and analysed, which can contribute to the development of adaptive parameter adjustment research. Originality/value The paper can be used to improve the effectiveness of ankle rehabilitation, to alleviate manual therapy problems in terms of labour intensiveness, precision and subjectivity and to ensure safety and comfort during rehabilitation sessions.
What problem does this paper attempt to address?