Design and control of a lower limb rehabilitation robot considering undesirable torques of the patient’s limb

Karam Almaghout,Bahram Tarvirdizadeh,Khalil Alipour,Alireza Hadi
DOI: https://doi.org/10.1177/0954411920947849
2020-08-11
Abstract:This research introduces a new exoskeleton-type rehabilitation robot, which can be used in lower limb rehabilitation therapy for post-stroke patients. A novel design of a typical knee and ankle rehabilitation robot is proposed. The kinematic and dynamic models of the knee and ankle rehabilitation robot are derived. Furthermore, a super-twisting nonsingular terminal sliding mode control is developed to achieve the desired training missions and its results are compared with those of an adaptive sliding mode control. To reduce undesired interaction torques between knee and ankle rehabilitation robot and patient, an admittance control algorithm is added to the controller to guarantee a safe therapy session. The admittance super-twisting nonsingular terminal sliding mode control structure is considered as the novelty of this article. Taking into account the dynamic uncertainties, external disturbances, and the interaction torques, the validity of the admittance super-twisting nonsingular terminal sliding mode control controller is approved by various numerical simulations over the admittance adaptive sliding mode control.
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to design and control a new type of lower - limb rehabilitation robot, especially for the knee and ankle joint rehabilitation training of post - stroke patients. Specifically, the paper focuses on the following points: 1. **Design a new rehabilitation robot**: The paper proposes a new exoskeleton - type rehabilitation robot design, which is intended for lower - limb rehabilitation treatment, especially for post - stroke patients. This robot can help patients perform highly repetitive task - specific training during the rehabilitation process to restore limb motor functions. 2. **Reduce unnecessary interaction torque**: During the rehabilitation training process, the patient's limbs may produce involuntary movements, and these movements will exert unnecessary interaction torque on the robot, which may cause injury to the patient. The paper proposes an impedance control algorithm, which reduces these unnecessary torques by adjusting the predefined trajectory, thereby ensuring the safety of the rehabilitation process. 3. **Development and comparison of control strategies**: In order to achieve accurate trajectory tracking and handle interaction torque, the paper develops two control strategies: - **Super - Twisting Nonsingular Terminal Sliding Mode Control (ST - NTSMC)**: This is an advanced sliding - mode control method, which can make the system state converge to the equilibrium point within a finite time, and has a fast response speed and high precision. - **Adaptive Sliding Mode Control (ASMC)**: This is a traditional sliding - mode control method for comparative study. 4. **System stability and robustness**: Through Lyapunov stability analysis, the paper proves that the proposed control strategies can ensure the stability and robustness of the system in the presence of dynamic uncertainties and external disturbances. 5. **Numerical simulation verification**: Through numerical simulation, the paper verifies the effectiveness of the proposed control strategies and compares them with the adaptive sliding - mode control, showing the superior performance of ST - NTSMC in dealing with dynamic uncertainties and external disturbances. In summary, this paper aims to improve the accuracy and safety of rehabilitation training by designing and controlling a new type of lower - limb rehabilitation robot, especially in dealing with the interaction torque generated by the patient's involuntary movements.