Comparison of bispectral index and patient state index during sevoflurane anesthesia in children: a prospective observational study

Sang-Hwan Ji,Young-Eun Jang,Eun-Hee Kim,Ji-Hyun Lee,Jin-Tae Kim,Hee-Soo Kim
DOI: https://doi.org/10.21203/rs.3.rs-53378/v1
2020-08-19
Abstract:Abstract Background: Cortical electroencephalography (EEG)-based devices are used to monitor the depth of anesthesia. In this study, we compared the values of bispectral index (BIS) and patient state index (PSI) during sevoflurane anesthesia in children. The ability/accuracy of BIS and PSI to predict the steady and recovery state of anesthesia was evaluated based on prediction probability (Pk) values and the secondary outcomes were agreement and correlation of two monitors. Methods: Fifty children (3-12 years old) were enrolled and the patients received sevoflurane anesthesia with remifentanil. Before the induction of anesthesia, BIS and PSI sensors were simultaneously placed on the forehead, and data were collected until the end of anesthesia. Steady state was defined as the period following intubation until the cessation of sevoflurane, while recovery state was defined as the period following the cessation of sevoflurane until awake. The prediction probability (Pk), agreement or correlation of BIS and PSI in different anesthesia state were calculated. Results: Anesthesia reduced mean BIS and PSI values. Pk of BIS (95% confidential interval (CI) [0.78-0.91])) and PSI (95% CI [0.82-0.91]) for anesthesia were 0.85 and 0.87, respectively. Agreement was 0.79 for recovery state and 0.73 for steady state. Conclusions: Pk values were comparable for BIS and PSI. Agreement between BIS and PSI measurements in the same state was relatively good. Therefore, these monitors are appropriate for monitoring for different state of anesthesia in pediatric population. Trial Registry Number: Clinical Trials.gov NCT03792334.
What problem does this paper attempt to address?