Evaluation of two automated low-cost RNA extraction protocols for SARS-CoV-2 detection
Fernando Lázaro-Perona,Carlos Rodriguez-Antolín,Marina Alguacil-Guillén,Almudena Gutiérrez-Arroyo,Jesús Mingorance,Julio García-Rodriguez,,
DOI: https://doi.org/10.1371/journal.pone.0246302
IF: 3.7
2021-02-16
PLoS ONE
Abstract:Background Two automatable in-house protocols for high-troughput RNA extraction from nasopharyngeal swabs for SARS-CoV-2 detection have been evaluated. Methods One hundred forty one SARS-CoV-2 positive samples were collected during a period of 10-days. In-house protocols were based on extraction with magnetic beads and designed to be used with either the Opentrons OT-2 (OT-2 in-house ) liquid handling robot or the MagMAX TM Express-96 system (MM in-house ). Both protocols were tested in parallel with a commercial kit that uses the MagMAX TM system (MM kit ). Nucleic acid extraction efficiencies were calculated from a SARS-CoV-2 DNA positive control. Results No significant differences were found between both in-house protocols and the commercial kit in their performance to detect positive samples. The MM kit was the most efficient although the MM in-house presented, in average, lower Cts than the other two. In-house protocols allowed to save between 350€ and 400€ for every 96 extracted samples compared to the commercial kit. Conclusion The protocols described harness the use of easily available reagents and an open-source liquid handling system and are suitable for SARS-CoV-2 detection in high throughput facilities.
multidisciplinary sciences
What problem does this paper attempt to address?