Infant heart-rate measurement and oxygen desaturation detection with a digital video camera using imaging photoplethysmography

Matthew E. Wieler,Thomas G. Murphy,Mira Blecherman,Hiral Mehta,G. Jesse Bender
DOI: https://doi.org/10.1038/s41372-021-00967-1
2021-03-01
Journal of Perinatology
Abstract:ObjectiveTo assess the feasibility of using an ordinary digital video camera to measure heart rate and detect oxygen desaturations in healthy infants.Study designHeart rate and oxygen saturation were measured with a video camera by detecting small color changes in 28 infants’ foreheads and compared with standard pulse oximetry measures. Multivariable regression examined the relationship between infant characteristics and heart-rate measurement precision.ResultsThe average bias of camera heart-rate measures was −4.2 beats per minute (BPM) and 95% limits of agreement were ±43.8 BPM. Desaturations detected by camera were 75% sensitive (15/20) and had a positive predictive value of 20% (15/74). Lower birth-weight was independently correlated with more precise heart-rate measures (8.05 BPM per kg, [95% CI 0.764–15.3]).ConclusionsA digital video camera provides accurate but imprecise measures of infant heart rate and may provide a rough screening tool for oxygen desaturations.
pediatrics,obstetrics & gynecology
What problem does this paper attempt to address?