Probing Heterogeneity in Attenuated Total Reflection Surface-Enhanced Infrared Absorption Spectroscopy (ATR-SEIRAS) Response with Synchrotron Infrared Microspectroscopy

Kaiyang Tu,Tyler A. Morhart,Stuart T. Read,Scott M. Rosendahl,Ian J. Burgess
DOI: https://doi.org/10.1177/00037028211005817
2021-04-07
Applied Spectroscopy
Abstract:The heterogeneity of metal island films electrodeposited on conductive metal oxide modified internal reflection elements is shown to provide a variable attenuated total reflection surface-enhanced infrared absorption spectroscopy (ATR-SEIRAS) response. A self-assembled monolayer of a ferrocene-terminated thiol monolayer (FcC 11 SH) was formed on the gold islands covering a single substrate, which was measured using both a conventional spectrometer and a custom-built horizontal microscope. Cyclic voltammetry and ATR-SEIRAS results reveal that the FcC 11 SH-modified substrate undergoes a reversible electron transfer and an associated re-orientation of both the ferrocene/ferrocenium headgroup and the hydrocarbon backbone. The magnitude of the absorption signal arising from the redox changes in the monolayer, as well as the IR signature arising from the ingress/egress of the perchlorate counterions, is shown to depend significantly on the size of the infrared beam spot when using a conventional Fourier transform infrared spectrometer. By performing equivalent measurements on a horizontal microscope, the primary cause of the differences in the signal level is found to be the heterogeneity in the density of gold islands on the conductive metal oxide.
spectroscopy,instruments & instrumentation
What problem does this paper attempt to address?