Woven Electroanalytical Biosensor for Nucleic Acid Amplification Tests

Shirin Khaliliazar,Ingrid Öberg Månsson,Andrew Piper,Liangqi Ouyang,Pedro Réu,Mahiar Max Hamedi
DOI: https://doi.org/10.1002/adhm.202100034
IF: 10
2021-04-30
Advanced Healthcare Materials
Abstract:Fiber‐based biosensors enable a new approach in analytical diagnostic devices. The majority of textile‐based biosensors, however, rely on colorimetric detection. Here a woven biosensor that integrates microfluidics structures in combination with an electroanalytical readout based on a thiol‐self‐assembled monolayer (SAM) for Nucleic Acid Amplification Testing, NAATs is shown. Two types of fiber‐based electrodes are systematically characterized: pure gold microwires (bond wire) and off‐the‐shelf plasma gold‐coated polyester multifilament threads to evaluate their potential to form SAMs on their surface and their electrochemical performance in woven textile. A woven electrochemical DNA (E‐DNA) sensor using a SAM‐based stem‐loop probe‐modified gold microwire is fabricated. These sensors can specifically detect unpurified, isothermally amplified genomic DNA of Staphylococcus epidermidis (10 copies/µL) by recombinase polymerase amplification (RPA). This work demonstrates that textile‐based biosensors have the potential for integrating and being employed as automated, sample‐to‐answer analytical devices for point‐of‐care (POC) diagnostics.
nanoscience & nanotechnology,engineering, biomedical,materials science, biomaterials
What problem does this paper attempt to address?