Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Profile

Madeline E. Carr,Kathryn E. Keenan,Robba Rai,Michael A. Boss,Peter Metcalfe,Amy Walker,Lois Holloway
DOI: https://doi.org/10.1002/mp.15645
IF: 4.506
2022-04-11
Medical Physics
Abstract:PurposeTo assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors which could impact the technical performance of ADC in the clinic were also completed, including changing the slice‐encoded imaging direction and the reference sample ADC value.MethodsADC acquisitions were performed monthly on an isotropic diffusion phantom over one year. Measurements of ADC %bias, coefficients of variation for short/long term repeatability and precision (CVST/CVLT and CVP), and b‐value dependency (Depb) were calculated. The measurements were then assessed according to the Quantitative Imaging Biomarker Alliance (QIBA) Diffusion Profile specifications.ResultsThe average of all measurements over the year were within Profile recommended ranges. This included when testing was performed in different imaging directions, and on samples that had different ADC reference values (0.4 – 1.1 μm2/ms). Results in the axial plane for the central water vial included a bias of + 0.05%, CVST /CVLT/CVP = 0.1%/ 0.9%/ 0.4% and Depb = 0.4%.ConclusionsThe technical performance of ADC on a radiotherapy dedicated MRI scanner over the course of 12 months was considered conformant to the QIBA Profile. Quantifying these metrics and factors that may affect the performance is essential in progressing the use of ADC clinically: ensuring that the observed change of ADC in a tissue is due to a physiological response and not measurement variability.This article is protected by copyright. All rights reserved
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?