Press-fit acetabular cup fixation: Principles and testing

W Macdonald,L V Carlsson,G J Charnley,C M Jacobsson
DOI: https://doi.org/10.1243/0954411991534780
1999-01-01
Abstract:Pre-clinical testing of the fixation of press-fit acetabular components of total hip prostheses relies on cadaver or synthetic bone, but the properties and geometry of bone models differ from those of physiological bone. Cup designs use varied mechanisms for initial stability in bone; therefore, using different analogues and tests is appropriate. Press-fit cup stability was tested in the following: firstly, polyurethane (PU) foam modelling can-cellous support; secondly, glass-fibre reinforced epoxide (GFRE) tubes modelling acetabular cortical support; thirdly, cadaveric acetabula. Three commercial cups [Harris-Galante II (H-G II), Zimmer; Optifix, Smith & Nephew, Richards; porous coated anatomic (PCA), Howmedica] and an experimental cup with enhanced rim fixation were tested in three modes: direct pull-out, lever-out and axial torque. The fixation stabilities measured in the PU and the GFRE models showed trends consistent with those in cadaver bone, differing in the oversizing and cup geometry. The experimental cup was significantly more secure in most modes than other cups; the H-G II and Optifix cups showed similar stabilities, lower than that of the experimental cup but greater than that of the PCA cup (analysis of variance and Tukey's highly significant difference test; p < 0.001). The stabilities measured in cadaver bone more closely approximated those in GFRE. The use of several bone analogues enables separation of fixation mechanisms, allowing more accurate prediction of in vivo performance.