Abstract:The assessment of measurement uncertainty of the scanning laser acoustic microscope (SLAM) has not been thoroughly evaluated experimentally. It is essential that the uncertainty of any measurement be fully understood. Both accuracy and precision were experimentally evaluated for the SLAM by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are ± 12% and ± 15%, respectively. And from Dew Coming 710, a silicon oil, the speed accuracy and precision are ± 2.9% and ± 0.4%, respectively. Further, an application of the assessment of precision was conducted using duplicate samples of dog skin and wound tissue. From these evaluations of a heterogeneous tissue, the estimated precision in the measurement of the attenuation coefficient and speed was ± 16% and ± 1.7%, respectively. [Work supported by NIH Grant AM 21557 and the Rehabilitation Research and Development Service of the Veterans' Administration.]