Improving deuterium metabolic imaging (DMI) signal‐to‐noise ratio by spectroscopic multi‐echo bSSFP: A pancreatic cancer investigation
Dana C. Peters,Stefan Markovic,Qingjia Bao,Dina Preise,Keren Sasson,Lilach Agemy,Avigdor Scherz,Lucio Frydman
DOI: https://doi.org/10.1002/mrm.28906
2021-06-30
Magnetic Resonance in Medicine
Abstract:<section class="article-section__content"><h3 class="article-section__sub-title section1"> Purpose</h3><p>Deuterium metabolic imaging (DMI) maps the uptake of deuterated precursors and their conversion into lactate and other markers of tumor metabolism. Even after leveraging <sup>2</sup>H's short T<sub>1</sub>s, DMI's signal-to-noise ratio (SNR) is limited. We hypothesize that a multi-echo balanced steady-state free precession (ME-bSSFP) approach would increase SNR compared to chemical shift imaging (CSI), while achieving spectral isolation of the metabolic precursors and products. </p></section><section class="article-section__content"><h3 class="article-section__sub-title section1"> Methods</h3><p>Suitably tuned <sup>2</sup>H ME-bSSFP (five echo times [TEs], ΔTE = 2.2 ms, repetition time [TR]/flip-angle = 12 ms/60°) was implemented at 15.2T and compared to CSI (TR/flip-angle = 95 ms/90°) regarding SNR and spectral isolation, in simulations, in deuterated phantoms and for the in vivo diagnosis of a mouse tumor model of pancreatic adenocarcinoma (N = 10). </p></section><section class="article-section__content"><h3 class="article-section__sub-title section1"> Results</h3><p>Simulations predicted an SNR increase vs. CSI of 3-5, and that the peaks of <sup>2</sup>H-water, <sup>2</sup>H<sub>6,6'</sub>-glucose, and <sup>2</sup>H<sub>3,3'</sub>-lactate can be well isolated by ME-bSSFP; phantoms confirmed this. In vivo, at equal spatial resolution (1.25 × 1.25 mm<sup>2</sup>) and scan time (10 min), <sup>2</sup>H<sub>6,6'</sub>-glucose's and <sup>2</sup>H<sub>3,3'</sub>-lactate's SNR were indeed higher for bSSFP than for CSI, three-fold for glucose (57 ± 30 vs. 19 ± 11, <i>P</i> < .001), doubled for water (13 ± 5 vs. 7 ± 3, <i>P</i> = .005). The time courses and overall localization of all metabolites agreed well, comparing CSI against ME-bSSFP. However, a clearer localization of glucose in kidneys and bladder, the detection of glucose-avid rims in certain tumors, and a heterogeneous pattern of intra-tumor lactate production could only be observed using ME-bSSFP's higher resolution. </p></section><section class="article-section__content"><h3 class="article-section__sub-title section1"> Conclusions</h3><p>ME-bSSFP provides greater SNR per unit time than CSI, providing for higher spatial resolution mapping of glucose uptake and lactate production in tumors.</p></section>
radiology, nuclear medicine & medical imaging