Construction and Evaluation of a Low Cost Interface for the Determination of Elemental Speciation by Gas Chromatography Coupled to Inductively Coupled Plasma Mass Spectrometry (GC‐ICP‐MS)

Michael J. Watts,Alex W. Kim,Daniel S. Vidler,Richard O. Jenkins,John F. Hall,Chris F. Harrington
DOI: https://doi.org/10.1080/10739140601000616
2007-02-01
Abstract:The construction and evaluation of a low cost, easily demountable interface to couple capillary gas chromatography to inductively coupled plasma mass spectrometry detection is described. Using this interface, the capillary column can be maintained at a high temperature through to the tip of the torch injector using a transfer line heated by a combination of hot argon and electrical resistance. The interface is suitable for analytes with boiling points up to 230°C, allowing for the analysis of low and high boiling analytes in a single injection. The system was evaluated by the determination of the butyltin species in a marine sediment CRM using conventional calibration with tripropyltin dichloride as the internal standard and the measurement of methylmercury in a tuna fish CRM via species‐specific isotope dilution analysis. Detailed information on the design and construction of the interface are included to facilitate its construction and use by other workers.
chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?