Recent developments in IBA analysis at CENTA, Bratislava

Miroslav Ješkovský,Jakub Kaizer,Ivan Kontuľ,Jakub Kvasniak,Ján Pánik,Jakub Zeman,Pavel P. Povinec
DOI: https://doi.org/10.1051/epjconf/202226101002
2022-01-01
EPJ Web of Conferences
Abstract:An experimental setup used at the CENTA (Centre for Nuclear and Accelerator Technologies) laboratory to carry out IBA (Ion Beam Analysis) techniques is described. PIXE (Particle Induced X-ray Emission) technique offers a unique way of non-destructive elemental analysis using accelerated ion beams. The calibration of instrumental constants (H-values) of SDD and BEGe detectors, used in the setup, is described in detail. The H-value was determined as a function of X-ray energy by using thin MicroMatter standards of chosen elements. The calibrated H-values were then used in GUPIXWIN for the determination of elemental concentrations in two IAEA reference materials, which were compared with reference values. A reasonable agreement was achieved between the measured and reference values, while the calibrated H-values will be used for further analyses.
What problem does this paper attempt to address?