Focused-ion-beam thinning of frozen-hydrated biological specimens for cryo-electron microscopy

Michael Marko,Chyongere Hsieh,Richard Schalek,Joachim Frank,Carmen Mannella
DOI: https://doi.org/10.1038/nmeth1014
IF: 48
2007-02-04
Nature Methods
Abstract:Cryo-electron microscopy can provide high-resolution structural information about cells and organelles in the nearly native, frozen-hydrated state. Applicability, however, is limited by difficulties encountered in preparing suitably thin, vitreously frozen biological specimens. We demonstrate, by cryo-electron tomography of Escherichia coli cells, that a focused ion beam (FIB) can be used to thin whole frozen-hydrated cells in a convenient and essentially artifact-free way.
biochemical research methods
What problem does this paper attempt to address?