Atomic Number and Electron Density Measurement Using a Conventional X-ray Tube and a CdTe Detector

Wenjuan Zou,Takuya Nakashima,Yoshiaki Onishi,Akifumi Koike,Bunji Shinomiya,Hisashi Morii,Yoichiro Neo,Hidenori Mimura,Toru Aoki
DOI: https://doi.org/10.1143/jjap.47.7317
IF: 1.5
2008-09-12
Japanese Journal of Applied Physics
Abstract:In order to apply the dual-energy technique to material identification, a new computed tomography scanning system was proposed using a conventional X-ray tube and a CdTe detector. This system can provide information of projection data at two distinct energy bands for scanned materials. After introducing an approximation, the measured projection data were reconstructed to obtain the distributions of the X-ray linear attenuation coefficients of the materials at two different energies. Then, the corresponding atomic number and electron density can be derived with the dual-energy X-ray computed tomography (DXCT) method adopted. By comparing the obtained results with theoretical ones, the feasibility of using this system for identifying low-Z materials was demonstrated in this study.