Safety and Function of a New Clinical Intracerebral Microinjection Instrument for Stem Cells and Therapeutics Examined in the Göttingen Minipig

Carsten R. Bjarkam,Andreas N. Glud,Lee Margolin,Karl Reinhart,Ron Franklin,Dorthe Deding,Kåre S. Ettrup,Lise M. Fitting,Mette S. Nielsen,Jens-Christian H. Sørensen,Miles G. Cunningham
DOI: https://doi.org/10.1159/000268743
IF: 1.643
2010-01-01
Stereotactic and Functional Neurosurgery
Abstract:BACKGROUND: A new intracerebral microinjection instrument (IMI) allowing multiple electrophysiologically guided microvolume injections from a single proximal injection path in rats has been adapted to clinical use by coupling the IMI to an FHC microTargeting Manual Drive, designed to be used with standard stereotactic frame-based systems and FHC frameless microTargeting Platforms.METHODS: The function and safety of the device was tested by conducting bilateral electrophysiologically guided microinjections of fluorescent microspheres in the substantia nigra of 4 Göttingen minipigs.RESULTS: The device was easy to handle and enabled accurate electrophysiologically guided targeting of the substantia nigra with minimal local tissue damage.CONCLUSION: The IMI is suitable for clinical use and may prove useful for various stereotactic procedures that require high levels of precision and/or three-dimensional distribution of therapeutics within the brain.
surgery,neurosciences,neuroimaging
What problem does this paper attempt to address?