DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF EDOXABAN TOSYLATE IN TABLET DOSAGE FORMS

B. Anupama,P. Tejaswi,KNV. Chenchu Lakshmi,A. Vishwanadh
DOI: https://doi.org/10.53879/id.57.07.11389
2020-10-08
INDIAN DRUGS
Abstract:A rapid, simple and precise reversed phase high performance liquid chromatography (RP-HPLC) method was developed for the estimation of edoxabantosylate in tablets. The quantification was carried out using a Phenomenex C-18 column (250×4.6 mm i.d., 5μm particle size) in isocratic mode with mobile phase comprising of ammonium acetate buffer andacetonitrile in the ratio of 50:50 (V/V) at a flow rate 1 mL/min. The eluent was monitored at 240 nm. The retention time of the drug was 3.486 min. The calibration curve was linear in the concentration range of 5-25 μg/mL and per cent recovery ranged from 98.25-101.6.The developed method was validated as per ICH guidelines and the results obtained were satisfactory.The method can be applied for routine quality control analysis of edoxabantosylate in tablet dosage forms.
What problem does this paper attempt to address?