Effects of the nasal passage on forced oscillation lung function measurements

Chuong Ngo,Karl Krüger,Thomas Vollmer,Stefan Winter,Bernhard Penzlin,Sylvia Lehmann,Steffen Leonhardt,Berno Misgeld
DOI: https://doi.org/10.1515/bmt-2016-0158
2017-11-27
Biomedical Engineering / Biomedizinische Technik
Abstract:Abstract The forced oscillation technique (FOT) is a non-invasive pulmonary function test which is based on the measurement of respiratory impedance. Recently, promising results were obtained by the application of FOT on patients with respiratory failure and obstructive sleep apnea (OSA). By using a nasal mask instead of a mouthpiece, the influences of the nasal passage and upper shunt alter the measured mechanical impedance. In this paper, we investigated the effects of the nasal passage and mask on FOT measurements from eight healthy subjects. A method for flow correction has been developed, which contains a pressure-flow characteristics compensation of the undetermined flow leakage at the face-mask interface. Impedance calculation and parameter estimation were performed in the frequency domain using fast Fourier transform (FFT). Average nasal parameters were
What problem does this paper attempt to address?