Using integrated correlative cryo-light and electron microscopy to directly observe syntaphilin-immobilized neuronal mitochondria in situ

Shengliu Wang,Shuoguo Li,Gang Ji,Xiaojun Huang,Fei Sun
DOI: https://doi.org/10.1007/s41048-017-0035-x
2017-03-21
Biophysics Reports
Abstract:Correlative cryo-fluorescence and cryo-electron microscopy (cryo-CLEM) system has been fast becoming a powerful technique with the advantage to allow the fluorescent labeling and direct visualization of the close-to-physiologic ultrastructure in cells at the same time, offering unique insights into the ultrastructure with specific cellular function. There have been various engineered ways to achieve cryo-CLEM including the commercial FEI iCorr system that integrates fluorescence microscope into the column of transmission electron microscope. In this study, we applied the approach of the cryo-CLEM-based iCorr to image the syntaphilin-immobilized neuronal mitochondria in situ to test the performance of the FEI iCorr system and determine its correlation accuracy. Our study revealed the various morphologies of syntaphilin-immobilized neuronal mitochondria that interact with microtubules and suggested that the cryo-CLEM procedure by the FEI iCorr system is suitable with a half micron-meter correlation accuracy to study the cellular organelles that have a discrete distribution and large size, e.g. mitochondrion, Golgi complex, lysosome, etc.
What problem does this paper attempt to address?