Encoding scheme design for gradient-free, nonlinear projection imaging using Bloch-Siegert RF spatial encoding in a low-field, open MRI system

Kartiga Selvaganesan,Yonghyun Ha,Heng Sun,Zhehong Zhang,Chenhao Sun,Anja Samardzija,Gigi Galiana,R. Todd Constable
DOI: https://doi.org/10.1038/s41598-024-53703-y
IF: 4.6
2024-02-09
Scientific Reports
Abstract:Eliminating conventional pulsed B 0 -gradient coils for magnetic resonance imaging (MRI) can significantly reduce the cost of and increase access to these devices. Phase shifts induced by the Bloch-Siegert shift effect have been proposed as a means for gradient-free, RF spatial encoding for low-field MR imaging. However, nonlinear phasor patterns like those generated from loop coils have not been systematically studied in the context of 2D spatial encoding. This work presents an optimization algorithm to select an efficient encoding trajectory among the nonlinear patterns achievable with a given hardware setup. Performance of encoding trajectories or projections was evaluated through simulated and experimental image reconstructions. Results show that the encodings schemes designed by this algorithm provide more efficient spatial encoding than comparison encoding sets, and the method produces images with the predicted spatial resolution and minimal artifacts. Overall, the work demonstrates the feasibility of performing 2D gradient-free, low-field imaging using the Bloch-Siegert shift which is an important step towards creating low-cost, point-of-care MR systems.
multidisciplinary sciences
What problem does this paper attempt to address?
The paper attempts to address the problem of achieving gradient-free nonlinear projection imaging in low-field open magnetic resonance imaging (MRI) systems. Specifically, the authors aim to design effective encoding schemes by utilizing Bloch-Siegert radiofrequency spatial encoding techniques to reduce the cost of MRI equipment and improve its accessibility. Traditional MRI systems rely on pulsed B0 gradient coils for spatial encoding, which not only increases the cost of the equipment but also leads to potential claustrophobia and high noise levels for patients during scanning. Therefore, this study explores a new approach by using phase shifts generated by the Bloch-Siegert effect for spatial encoding, thereby achieving gradient-free, low-cost, open MRI imaging. The focus of the research is on developing an optimization algorithm to select effective encoding trajectories that can be realized under given hardware settings, particularly in 2D spatial encoding. The performance of these encoding schemes was evaluated through simulated and experimental image reconstructions. The results show that the encoding schemes designed using this method provide more efficient spatial encoding compared to other encoding sets, capable of producing images with the desired spatial resolution and minimal artifacts. This marks an important step towards creating low-cost, portable MRI systems.