High-intensity focused ultrasound therapy in the uterine fibroid: a clinical case study of poor heating efficacy

Visa Suomi,Antti Viitala,Teija Sainio,Gaber Komar,Roberto Blanco Sequeiros
DOI: https://doi.org/10.48550/arXiv.1910.12211
IF: 4.506
2019-10-27
Medical Physics
Abstract:A clinical case study of high-intensity focused ultrasound (HIFU) treatment in the uterine fibroid was conducted. During the therapy, poor heating efficacy was observed which could be attributed to several factors such as the local perfusion rate, patient-specific anatomy or changes in acoustic parameters of the ultrasound field. In order to determine the cause of the diminished heating, perfusion analyses and ultrasound simulations were conducted using the magnetic resonance imaging (MRI) data from the treatment. The perfusion analysis showed high local perfusion rate in the myoma (301.0 +- 25.6 mL/100 g/min) compared to the surrounding myometrium (233.8 +- 16.2 mL/100 g/min). The ultrasound simulations did not show large differences in the focal point shape or the acoustic pressure (2.07 +- 0.06 MPa) when tilting the transducer. However, a small shift (-2.2 +- 1.3 mm) in the axial location of the focal point was observed. The main causes for the diminished heating were likely the high local perfusion and ultrasound attenuation due to the deep location of the myoma.