Development of an on-site therapeutic drug monitoring method using a portable spectrometer

Kenichiro Todoroki,Nan Fukudo,Yuto Kudoh,Hajime Mizuno,Jun Zhe Min,Shimako Tanaka,Shinya Uchida,Noriyuki Namiki,Toshimasa Toyo’oka
DOI: https://doi.org/10.1007/s44211-024-00513-x
IF: 1.967
2024-02-16
Analytical Sciences
Abstract:We report on the development of an on-site therapeutic drug monitoring (TDM) method for vancomycin (VCM) utilizing a portable spectrometer and commercially available immunoturbidimetric assay reagents designed for automated clinical chemistry analyzers. The method enables the quantification of VCM in plasma samples within 10 min, with a good correlation between the measured values and the theoretical values ( r 2 = 0.995). The intra and inter-day precisions were found to be below 12.5% and 17.7%, respectively. Moreover, we established a correlation between the quantitative values using this method and those measured through HPLC–UV and automated clinical chemistry analyzers, showing good reliability ( R 2 = 0.970 and 0.951, respectively). This method allows anyone to rapidly perform TDM at the bedside and is expected to be used to evaluate appropriate drug therapy.
chemistry, analytical
What problem does this paper attempt to address?