Abstract:Journal of Hand Surgery (European Volume), Volume 49, Issue 2, Page 177-187, February 2024. We reviewed the incidence and management of complications after total wrist arthroplasty, as reported in the literature, with so-called fourth-generation implants and other recent designs. While early intraoperative and postoperative complications, including fractures, tendon lacerations, infection, nerve compression, tendonitis, stiffness and chronic regional pain syndrome, had an acceptable incidence, late complications, such as periprosthetic osteolysis and implant loosening, occurred more frequently. Implant survival at 10 years was in the range of 70%–80% in most publications. Several of the implants have been modified or withdrawn. Instability and dislocation were frequent after a pyrocarbon spacer. Failed arthroplasties can be salvaged by revision arthroplasty or total wrist arthrodesis. Revision arthroplasty has a lower survival rate than primary arthroplasty and does not clearly offer important significant advantages over total wrist arthrodesis in terms of patient-reported outcome measures. Further development of prosthetic design, new materials and more knowledge on patient-related risk factors are needed.