Reconstruction algorithm for frequency-differential EIT using absolute values

Tobias Menden,Jakob Orschulik,Sébastien Dambrun,Jascha Matuszczyk,Susana Aguiar Santos,Steffen Leonhardt,Marian Walter
DOI: https://doi.org/10.1088/1361-6579/ab0b55
2019-04-03
Physiological Measurement
Abstract:Objective : Tissues in the body differ by their frequency-dependent conductivity.Frequency-differential electrical impedance tomography (fdEIT) is a promising technique toreconstruct the distribution of tissue inside the body by injecting current at two frequencies andmeasuring the resulting surface-potential. Approach : The Gauss–Newton method is one way to map thesurface measurements to a conductivity image. Usually, the minimization function contains onlyweighted differential measurement data and a regularization. This traditional method is extended byabsolute measurement data to improve fdEIT reconstruction results. The key challenge of unknowntorso geometries and electrode displacement has been addressed for the reconstruction of differentlung pathologies. Main results : The frequency-dependent conductivity of the background wasreconstructed precisely and a contrast between organs was achieved. The algorithm shows goodperformance compared to GRE...
engineering, biomedical,biophysics,physiology
What problem does this paper attempt to address?