Gated tomographic radionuclide angiography using cadmium-zinc-telluride detector gamma camera; comparison to traditional gamma cameras

Maria Maj Jensen,Ulla Schmidt,Chenxi Huang,Bo Zerahn
DOI: https://doi.org/10.1007/s12350-013-9844-6
2013-12-24
Journal of Nuclear Cardiology
Abstract:PurposeEstimation of left ventricular ejection fraction (LVEF) with equilibrium 99MTc-HSA equilibrium radionuclide angiography (MUGA) is frequently used for assessing cardiac function. The purpose of this study was to compare intra- and interobserver variation between three different gamma cameras.Materials and MethodsEighty-two patients, scanned in the same sequential order on the three cameras. Each acquisition was analyzed twice by two technologists. Inter- and intraobserver variations were calculated as the coefficient of variation and the 95% confidence interval for limits of agreement between each sequence of analyses for each of the three cameras.ResultsThe lowest intraobserver variations in LVEF for the two NaI-detector cameras were 3.1% (−4.0% to 3.5%) for the planar and 3.4% (−4.2% to 4.5%) for SPECT (P ≤ 0.001-0.019), the highest result for the CZT SPECT camera was 2.6% (−2.9% to 3.1%). Similarly, interobserver variation was 4.8% (−4.8% to 6.4%) and 4.9% (−5.4% to 7.5%), respectively, for each of the NaI-detector cameras and 3.3% (−3.4% to 4.3%) for the CZT SPECT camera (P ≤ 0.001-0.008).DiscussionThe CZT detector camera was superior to both NaI detector cameras regarding intra- and interobserver variation. The CZT SPECT camera may identify changes in LVEF with greater certainty than its NaI detector-equipped counterparts.
radiology, nuclear medicine & medical imaging,cardiac & cardiovascular systems
What problem does this paper attempt to address?