Effects of Blood Collection Tubes on the Quantification of Catecholamines and Their O‐Methylated Metabolites by Liquid Chromatography Coupled With Tandem Mass Spectrometry

Rong Su,Yanjie Wu,Qiuyi Huang,Shengchun Huang,Yao Li,Caifeng Chen,Wenchun Xie,Yan Li
DOI: https://doi.org/10.1002/jcla.25107
IF: 3.124
2024-10-02
Journal of Clinical Laboratory Analysis
Abstract:The effects of blood collection tubes on the analysis of catecholamines and metanephrines by LC‐MS/MS were not clear. In this study, we compared the retention behavior and assay performance of catecholamines and metanephrines in serum, heparin, and K2EDTA plasma and found that the results for catecholamines were more affected by the collection tubes, especially for norepinephrine. We suggest that it is necessary to evaluate the blood collection tubes before the test are performed in clinical laboratory. Background Catecholamines (epinephrine; norepinephrine; and dopamine) and their O‐methylated metabolites (metanephrine; normetanephrine; and 3‐methoxytyramine) are biomarkers for pheochromocytoma and paraganglioma. Liquid chromatography coupled with tandem mass spectrometry (LC‐MS/MS) was recommended by Endocrine Society for detecting these compounds. The influence of blood collection tubes on the analysis of the six analytes by LC‐MS/MS was not thoroughly investigated, which we want to clarify in our study. Methods Blood samples of healthy individuals were collected into serum, lithium heparin, and K2EDTA plasma tubes separately. Samples were subjected to solid phase extraction and then analyzed by LC‐MS/MS. The retention behavior and assay performance of the six analytes were assessed for samples from different collection containers. The impacts of potassium and sodium as the counter ions of EDTA on the retention time and matrix effect were also studied. Results Compared with O‐methylated metabolites, the results for catecholamines were more affected by the collection tubes, especially for norepinephrine, which displayed severely suppressed signal and very low extraction efficiency in K2EDTA plasma. Changing the counter ion of EDTA from potassium to sodium dramatically changed the retention behavior and matrix effect of norepinephrine. Conclusions It is necessary to evaluate blood collection tubes for catecholamines and their O‐methylated metabolites analyzed by LC‐MS/MS. In addition, attention should also be paid when the anticoagulant counter ion was changed.
medical laboratory technology
What problem does this paper attempt to address?