Rapid quantitative detection of glucose content in glucose injection by reaction headspace gas chromatography

Wei-Qi Xie,Yi-Xian Gong,Kong-Xian Yu
DOI: https://doi.org/10.1016/j.chroma.2017.09.018
IF: 4.601
2017-10-01
Journal of Chromatography A
Abstract:This work investigates an automated technique for rapid detecting the glucose content in glucose injection by reaction headspace gas chromatography (HS-GC). This method is based on the oxidation reaction of glucose in glucose injection with potassium dichromate. The carbon dioxide (CO<sub>2</sub>) formed from the oxidation reaction can be quantitatively detected by GC. The results show that the relative standard deviation (RSD) of the present method was within 2.91%, and the measured glucose contents in glucose injection closely match those quantified by the reference method (relative differences &lt;6.45%). The new HS-GC technique is rapid, practical and can be used to the batch detection of the glucose content in glucose injection related applications.
chemistry, analytical,biochemical research methods
What problem does this paper attempt to address?