Implantable bioelectronic systems for early detection of kidney transplant rejection

Surabhi R Madhvapathy,Jiao-Jing Wang,Heling Wang,Manish Patel,Anthony Chang,Xin Zheng,Yonggang Huang,Zheng J Zhang,Lorenzo Gallon,John A Rogers,Surabhi R. Madhvapathy,Zheng J. Zhang,John A. Rogers
DOI: https://doi.org/10.1126/science.adh7726
IF: 56.9
2023-09-08
Science
Abstract:Early-stage organ transplant rejection can be difficult to detect. Percutaneous biopsies occur infrequently and are risky, and measuring biomarker levels in blood can lead to false-negative and -positive outcomes. We developed an implantable bioelectronic system capable of continuous, real-time, long-term monitoring of the local temperature and thermal conductivity of a kidney for detecting inflammatory processes associated with graft rejection, as demonstrated in rat models. The system detects ultradian rhythms, disruption of the circadian cycle, and/or a rise in kidney temperature. These provide warning signs of acute kidney transplant rejection that precede changes in blood serum creatinine/urea nitrogen by 2 to 3 weeks and approximately 3 days for cases of discontinued and absent administration of immunosuppressive therapy, respectively.
multidisciplinary sciences
What problem does this paper attempt to address?