Adequacy of an in-NICU 1T MRI Compared to 3T MRI for Clinical Management

Danielle Sharon,Elizabeth Singh,Camilo Jaimes,Ellen Grant,Terrie Inder,Mohamed El-Dib
DOI: https://doi.org/10.1016/j.pediatrneurol.2024.08.004
IF: 4.21
2024-08-18
Pediatric Neurology
Abstract:This retrospective study aims to assess the added diagnostic utility and clinical value of a 3-Tesla neonatal brain Magnetic Resonance Imaging after obtaining a 1-Tesla Magnetic Resonance Imaging within the Neonatal Intensive Care Unit. A cohort of 34 infants had an initial 1-Tesla Magnetic Resonance Imaging and repeat imaging within 14 days in a 3-Tesla scanner. All infants were admitted to the level-III Neonatal Intensive Care Unit at Brigham and Women's Hospital and all images were interpreted by pediatric neuroradiologists. In 31 cases (91%), the 3-Tesla Magnetic Resonance Imaging showed similar or expected evolution of known findings found on 1-Tesla. In cases with change between the 1-Tesla and 3-Tesla imaging results, there was no clinical impact. In conclusion, Images from 1-Tesla Magnetic Resonance Imaging were sufficient for characterizing a wide range of neonatal brain injuries and abnormalities and repeated 3-Tesla Magnetic Resonance Imaging did not yield further clinical benefit.