A Setup for Direct Measurement of the Adiabatic Temperature Change in Magnetocaloric Materials
Sagar Ghorai,Daniel Hedlund,Martin Kapuscinski,Peter Svedlindh
DOI: https://doi.org/10.1109/tim.2023.3272387
IF: 5.6
2023-05-19
IEEE Transactions on Instrumentation and Measurement
Abstract:In order to find a highly efficient, environmentally-friendly magnetic refrigerant, direct measurements of the adiabatic temperature change are required. Here, in this work, a simple setup for the measurement is presented. Using a permanent magnet Halbach array with a maximum magnetic field of 1.8 T and a rate of magnetic field change of 5 T/s, accurate determination of is possible in this system. The operating temperature range of the system is from 100 to 400 K, designed for the characterization of materials with potential for room temperature magnetic refrigeration applications. Using the setup, of a first-order and two second-order compounds have been studied. Results from the direct measurement for the first-order compound have been compared with calculated from the temperature and magnetic field-dependent specific heat data. By comparing results from direct and indirect measurements, it is concluded that for a reliable characterization of the magnetocaloric effect (MCE), direct measurement of should be adopted.
engineering, electrical & electronic,instruments & instrumentation