Iterative Reconstruction-Reprojection: An Algorithm for Limited Data Cardiac-Computed Tomography

Menahem Nassi,William R. Brody,Barry P. Medoff,Albert Macovski
DOI: https://doi.org/10.1109/tbme.1982.324900
IF: 4.6
1982-05-01
IEEE Transactions on Biomedical Engineering
Abstract:Cardiac X-ray computed tomography (CT) has been limited due to scanning times which are considerably longer (1 s) than required to resolve the beating heart (0.1 s). The otherwise attractive convolution-backprojection algorithm is not suited for CT image reconstruction from measurements comprising an incomplete set of projection data. In this paper, an iterative reconstruction-reprojection (IRR) algorithm is proposed for limited projection data CT image reconstruction. At each iteration, the missing views are estimated based on reprojection, which is a software substitute for the scanning process. The standard fan-beam convolution-backprojection algorithm is then used for image reconstruction. The proposed IRR algorithm enables the use of convolution-backprojection in limited angle of view and in limited field of view CT cases. The potential of this method for cardiac CT reconstruction is demonstrated using computer simulated data.
engineering, biomedical
What problem does this paper attempt to address?