Imaging of 17 O-labeled Water Using Fast T2 Mapping with T2-preparation: A Phantom Study

Hiroyuki Kameda,Yumi Nakada,Yuta Urushibata,Hiroyuki Sugimori,Takaaki Fujii,Naoya Kinota,Daisuke Kato,Minghui Tang,Keita Sakamoto,Kohsuke Kudo
DOI: https://doi.org/10.2463/mrms.tn.2023-0152
2024-01-01
Magnetic Resonance in Medical Sciences
Abstract:<sup>17</sup>O-labeled water is a T2-shortening contrast agent used in proton MRI and is a promising method for visualizing cerebrospinal fluid (CSF) dynamics because it provides long-term tracking of water molecules. However, various external factors reduce the accuracy of <sup>17</sup>O-concentration measurements using conventional signal-intensity-based methods. In addition, T2 mapping, which is expected to provide a stable assessment, is generally limited to temporal-spatial resolution. We developed the T2-prepared based on T2 mapping used in cardiac imaging to adapt to long T2 values and tested whether it could accurately measure <sup>17</sup>O-concentration in the CSF using a phantom. The results showed that <sup>17</sup>O-concentration in a fluid mimicking CSF could be evaluated with an accuracy comparable to conventional T2-mapping (Carr-Purcell-Meiboom-Gill multi-echo spin-echo method). This method allows <sup>17</sup>O-imaging with a high temporal resolution and stability in proton MRI. This imaging technique may be promising for visualizing CSF dynamics using <sup>17</sup>O-labeled water.