Vector-flow imaging of slowly moving ex vivo blood with photoacoustics and pulse-echo ultrasound

Caitlin Smith,Jami Shepherd,Guillaume Renaud,Kasper van Wijk
DOI: https://doi.org/10.1016/j.pacs.2024.100602
IF: 9.656
2024-03-16
Photoacoustics
Abstract:We present a technique called photoacoustic vector-flow (PAVF) to quantify the speed and direction of flowing optical absorbers at each pixel in an acoustic-resolution photoacoustic image. By varying the receiving angle at each pixel in post-processing, we obtain multiple estimates of the phase difference between consecutive frames. These are used to solve the overdetermined photoacoustic Doppler equation with a least-squares approach to estimate a velocity vector at each pixel. This technique is tested in bench-top experiments and compared to simultaneous pulse-echo ultrasound vector-flow (USVF) on whole rat blood at speeds on the order of 1 mm/s. Unlike USVF, PAVF can detect flow without stationary clutter filtering in this experiment, although the velocity estimates are highly underestimated. When applying spatio-temporal singular value decomposition clutter filtering, the flow speed can be accurately estimated with an error of 17% for USVF and 9% for PAVF for an average flow speed of 2.5 mm/s.
engineering, biomedical,radiology, nuclear medicine & medical imaging,instruments & instrumentation