Study of the single cluster response of a helium-isobutane drift chamber prototype using 8 keV X-rays

G. Cavoto,S. Dabagov,D. Hampai,G. Piredda,F. Renga,E. Ripiccini,C. Voena,A. Zullo
DOI: https://doi.org/10.1088/1748-0221/10/03/p03012
2015-03-11
Journal of Instrumentation
Abstract:The identification of single clusters in the electronic signals produced by ionizing particles within a drift chamber is expected to significantly improve the performances of this kind of detectors in terms of particle identification capabilities and space resolution. In order to develop refined cluster recognition algorithms, it is essential to measure the response of the chamber and its electronics to single ionization clusters. This can be done by irradiating the chamber with X-rays. We report here on the studies performed on a drift chamber prototype for the MEG-II experiment at the X-ray facility of the INFN Frascati's National Laboratories ``XLab Frascati''. The prototype is operated with a helium-isobutane mixture and instrumented with high bandwidth custom pre-amplifiers. The results of this study have been used to develop an innovative method for cluster recognition, based on the Wiener filter technique, which has been tested on data collected at the Frascati's Beam Test Facility. As a side measurement, we also performed a study of the gas gain in a configuration which is similar to that of the MEG-II experiment.
instruments & instrumentation
What problem does this paper attempt to address?