A Fully Automated Segmentation and Morphometric Parameter Estimation System for Assessing Corneal Endothelial Cell Images

Jing-Hao Qu,Xiao-Ran Qin,Rong-Mei Peng,Ge-Ge Xiao,Jian Cheng,Shao-Feng Gu,Hai-Kun Wang,Jing Hong
DOI: https://doi.org/10.1016/j.ajo.2022.02.026
IF: 5.488
2022-07-01
American Journal of Ophthalmology
Abstract:Background: The goal of this study was to develop a fully automated segmentation and morphometric parameter estimation system for assessing corneal endothelial cells from in vivo confocal microscopy (IVCM) images.Design: Artificial intelligence (neural network) study.Methods: First, we developed a fully automated deep learning system for assessing corneal endothelial cells using the development set (from ninety-nine subjects). Second, 184 images (from ninety-seven subjects) were used to construct the testing set to evaluate the clinical validity and usefulness of the automated segmentation and morphometric system. Third, the automatically calculated endothelial cell density (ECD) values, Topcon's cell density and the manually calculated ECD were compared.Results: After slit lamp examination, we identified eighty-eight healthy subjects, two Fuchs endothelial dystrophy patients and seven corneal endotheliitis patients among the ninety-seven subjects in the testing set. The automatedly estimated morphometric parameters for the testing set were an average number of cells of 234, an ECD of 2592 cells/mm2, a coefficient of variation in the cell area (CV) of 32.14% and a percentage of hexagonal cells (HEX) of 54.16%. The Pearson's correlation coefficient between the automated ECD and Topcon's cell density and between the manually calculated ECD and Topcon's cell density was 0.932 (P<0.01) and 0.818 (P<0.01), respectively. The Bland–Altman plot of Topcon's cell density and the automated ECD yielded 95% limits of agreement between 271.94 and -572.46 (concordance correlation coefficient = 0.9).Conclusions: A fully automated method for segmenting corneal endothelial cells and estimating morphometric parameters using IVCM images is more efficient and accurate for assessing the normal corneal endothelium.
ophthalmology
What problem does this paper attempt to address?