Development, validation and application of an ion-pair reversed-phase liquid chromatography-tandem mass spectrometry method for the quantification of nusinersen

Xiao Zhang,Chunjie Sha,Wei Zhang,Fengjuan Zhao,Mingli Zhu,Guangyi Leng,Wanhui Liu
DOI: https://doi.org/10.4155/bio-2023-0218
IF: 2.695
2024-02-11
Bioanalysis
Abstract:Background: The fully phosphorothioate-modified oligonucleotide (OGN) nusinersen has low ionization efficiency in the negative ion mode, resulting in a low mass spectrometry response. There have been no relevant reports on developing a LC–MS method for the determination of nusinersen by optimizing mobile phase composition. Materials & methods: Mobile phase additives comprised of 15 mM triethylamine/25 mM 1,1,1,3,3,3-hexafluoro-2-propanol with a pH of 9.6. Nusinersen was extracted from plasma using Oasis ® HLB solid-phase extraction (Waters, MA, USA). Results & conclusion: By adjusting the pH of the mobile phase to 9.6 by optimizing the type and concentration of ion-pair reagents, a high mass spectrometry response was obtained. The developed method was applied to nusinersen and met the requirements for the pharmacokinetic study of nusinersen in rabbits.
biochemical research methods,chemistry, analytical
What problem does this paper attempt to address?