Rapid removal of IgG1 carryover on protease column using protease-safe wash solutions delivered with LC pump for HDX-MS systems

Kyle Anderson,Sandeep Kumar
DOI: https://doi.org/10.26434/chemrxiv-2024-8vg7n
2024-10-21
Abstract:Sample carryover is a common problem in hydrogen-deuterium exchange mass spectrometry, particularly because immobilized protease columns cannot withstand the high organic solvent concentrations typically used in liquid chromatography mass spectrometry (LC-MS) for cleaning. Conventional cleaning methods using injections of guanidine HCl still suffer from carryover and may require four blanks after each sample run to fully remove carryover. We have implemented an additional LC pump to deliver customized wash solutions to protease and enzyme columns and the associated LC capillaries to eliminate carryover. Pump-based washing using the protease-safe wash solutions tested herein were able to fully remove carryover with only one blank run. FOS-choline-12 was found to be the most effective component in wash solutions and even performed well alone at 0.1 % volume concentration. Since the protease column washing is performed concurrently during the analytical gradient within a sample run, subsequent blank runs could be reduced from four to one and total run time could be reduced by up to 60 %. Savings in total run time could more than double the productivity of data acquisition, which is imperative for pandemic preparedness and for acceleration of biotherapeutics development.
Chemistry
What problem does this paper attempt to address?
This paper attempts to address the common issue of sample carryover in hydrogen-deuterium exchange mass spectrometry (HDX-MS). Specifically, due to the immobilized protease column's intolerance to the high organic solvent concentrations typically used for liquid chromatography-mass spectrometry (LC-MS) cleaning, traditional cleaning methods such as injecting guanidine hydrochloride solution still face sample carryover issues and require multiple blank runs to completely remove the residue. This not only increases experimental time but also reduces data acquisition efficiency. To solve this problem, the authors designed an additional liquid chromatography pump to deliver a customized cleaning solution to the protease column and associated liquid chromatography capillaries to thoroughly eliminate sample carryover. Through this method, the authors found that using a mild cleaning solution containing 0.1% FOS-choline-12 and 0.1% formic acid, only one blank run was needed to completely remove the residue, thereby reducing the total run time by 60% and significantly improving data acquisition efficiency. This improvement is of great significance for pandemic preparedness and accelerating the development of biotherapeutic drugs.