Automatic elasticity measurement of single cells using a microfluidic system with real-time image processing*
Siyuan Chen,Tianruo Guo,Jing Jin,Dong Xu,Huaying Chen,Yike Cai
DOI: https://doi.org/10.1109/EMBC40787.2023.10340799
2023-07-01
Abstract:The mechanical properties of cells are closely related to their physiological states and functions. Due to the limitations of conventional cell elasticity measurement technologies such as low throughput, cell-invasiveness, and high cost, microfluidic systems are emerging as powerful tools for high-throughput cell mechanical property studies. This paper introduces a microfluidic system to automatically measure the elastic modulus of single cells in real time. The system integrated a microfluidic chip with a microchannel for cell constriction, a pressure pump, a precision differential pressure sensor, and a program for online analysis of cell deformation. The program used a fast U-net to segment cell images and measure protrusion length during cell deformation. Subsequently, the cell elasticity was determined in real-time based on the deformation and required pressure using the power law rheological model. Finally, Young's modulus of BMSCs, Huh-7 cells, EMSCs, and K562 cells was measured as 25.13 ± 15.19 Pa, 69.74 ± 92.01 Pa, 54.50 ± 59.31 Pa and 58.43 ± 27.27 Pa, respectively. The microfluidic system has significant application potential in the automated evaluation of cell mechanical properties.Clinical Relevance—The technique in this paper may be used for the automatic and high throughput study of the stiffness of cells, such as stem cells and cancer cells. The stiffness data may contribute to stem cell therapy and cancer research.
Medicine,Computer Science,Engineering