Exploring the Potential of a New Wearable Sleep Monitoring Device for Clinical Application

Xin Li,Min Li,Mei Tian,Qi Liu,Xiaomeng Zhou,Hu Liu,Rui Li,Zhenzhong Li,Hui Dong,Lijing Jia,Yaling Liu
DOI: https://doi.org/10.1016/j.bspc.2024.106856
IF: 5.1
2025-01-01
Biomedical Signal Processing and Control
Abstract:OBJECTIVES Sleep is critical to health and quality of life, and the gold-standard assessment method for sleep disorders--polysomnography (PSG)--has struggled to become widely used because of its inherent shortcomings, which portable sleep-monitoring devices are expected to improve. This study aims to validate a new EEG-, EOG-, and EMG-based portable sleep-monitoring device against conventional PSG to explore its feasibility for clinical use. Methods Twenty participants completed overnight sleep monitoring using conventional polysomnography and the Wearable PolysomnogramTM (WPSG-I). The similarity of the raw signals recorded by the two devices and the agreement rate of their sleep-staging results were compared. Results The raw electroencephalogram (EEG) signals recorded by the two devices were highly similar, and the agreement rates between WPSG-I manual sleep staging and automatic sleep staging with conventional PSG manual sleep staging were 95.8 % (kappa = 0.92) and 89.7 % (kappa = 0.80), respectively. Conclusion The agreement rate of sleep staging results between the two devices was high. The results of this study demonstrate the validity and feasibility of the WPSG-I for sleep research and clinical applications.
What problem does this paper attempt to address?